LibTiePie  0.9.11
Library for interfacing TiePie engineering instruments

Description

Functions to control the probe settings of an oscilloscope channel.

Probes connected to an input channel usually attenuate the input signal. Specialized probes or signal converters can also amplify the input signal and/or apply an offset to it. Probe gain and offset settings are available to compensate the measured values for the gain and offset applied by the probe or converter. The channel will then give the original signal values.

When e.g. a X10 attenuating oscilloscope probe is connected, the channel will measure a 10 times smaller value of the input signal. By setting the probe gain to 10, the measured values will be multiplied by 10 to compensate.

By default probe gain is set to 1x and probe offset to 0.

Functions

double ScpChGetProbeGain (LibTiePieHandle_t hDevice, uint16_t wCh)
 Get the currently set channel probe gain for a specified channel. More...
 
double ScpChSetProbeGain (LibTiePieHandle_t hDevice, uint16_t wCh, double dProbeGain)
 Set the channel probe gain for a specified channel. More...
 
double ScpChGetProbeOffset (LibTiePieHandle_t hDevice, uint16_t wCh)
 Get the currently set channel probe offset for a specified channel. More...
 
double ScpChSetProbeOffset (LibTiePieHandle_t hDevice, uint16_t wCh, double dProbeOffset)
 Set the channel probe offset for a specified channel. More...
 

Function Documentation

double ScpChGetProbeGain ( LibTiePieHandle_t  hDevice,
uint16_t  wCh 
)

Get the currently set channel probe gain for a specified channel.

Parameters
[in]hDeviceA device handle identifying the oscilloscope.
[in]wChThe channel number identifying the channel, 0 to ScpGetChannelCount() - 1.
Returns
The currently set probe gain.
Status values
INVALID_CHANNEL The requested channel number is invalid.
INVALID_HANDLE The handle is not a valid oscilloscope handle.
OBJECT_GONE The object indicated by the handle is no longer available.
LIBRARY_NOT_INITIALIZEDThe library is not initialized, see LibInit().
SUCCESS The function executed successfully.
See also
ScpChSetProbeGain
ScpChGetDataValueRange
ScpChGetDataValueMin
ScpChGetDataValueMax
Since
0.4.0
double ScpChSetProbeGain ( LibTiePieHandle_t  hDevice,
uint16_t  wCh,
double  dProbeGain 
)

Set the channel probe gain for a specified channel.

Parameters
[in]hDeviceA device handle identifying the oscilloscope.
[in]wChThe channel number identifying the channel, 0 to ScpGetChannelCount() - 1.
[in]dProbeGainThe required probe gain, -1e6 <= gain <= 1e6, != 0.
Returns
The actually set probe gain.
Status values
INVALID_VALUE The requested probe gain value is invalid.
VALUE_CLIPPED The requested probe gain value is outside the valid range and clipped to the closest limit.
INVALID_CHANNEL The requested channel number is invalid.
INVALID_HANDLE The handle is not a valid oscilloscope handle.
OBJECT_GONE The object indicated by the handle is no longer available.
LIBRARY_NOT_INITIALIZEDThe library is not initialized, see LibInit().
SUCCESS The function executed successfully.
See also
ScpChGetProbeGain
ScpChGetDataValueRange
ScpChGetDataValueMin
ScpChGetDataValueMax
Since
0.4.0
double ScpChGetProbeOffset ( LibTiePieHandle_t  hDevice,
uint16_t  wCh 
)

Get the currently set channel probe offset for a specified channel.

The probe offset value is in Volt, Ampere or Ohm, depending on the input coupling

Parameters
[in]hDeviceA device handle identifying the oscilloscope.
[in]wChThe channel number identifying the channel, 0 to ScpGetChannelCount() - 1.
Returns
The currently set probe offset.
Status values
INVALID_CHANNEL The requested channel number is invalid.
INVALID_HANDLE The handle is not a valid oscilloscope handle.
OBJECT_GONE The object indicated by the handle is no longer available.
LIBRARY_NOT_INITIALIZEDThe library is not initialized, see LibInit().
SUCCESS The function executed successfully.
See also
ScpChSetProbeOffset
ScpChGetDataValueRange
ScpChGetDataValueMin
ScpChGetDataValueMax
Since
0.4.4
double ScpChSetProbeOffset ( LibTiePieHandle_t  hDevice,
uint16_t  wCh,
double  dProbeOffset 
)

Set the channel probe offset for a specified channel.

The probe offset value is in Volt, Ampere or Ohm, depending on the input coupling

Parameters
[in]hDeviceA device handle identifying the oscilloscope.
[in]wChThe channel number identifying the channel, 0 to ScpGetChannelCount() - 1.
[in]dProbeOffsetThe required probe offset, -1e6 <= offset <= 1e6.
Returns
The actually set probe offset.
Status values
VALUE_CLIPPED The requested probe offset value is outside the valid range and clipped to the closest limit.
INVALID_CHANNEL The requested channel number is invalid.
INVALID_HANDLE The handle is not a valid oscilloscope handle.
OBJECT_GONE The object indicated by the handle is no longer available.
LIBRARY_NOT_INITIALIZEDThe library is not initialized, see LibInit().
SUCCESS The function executed successfully.
See also
ScpChGetProbeOffset
ScpChGetDataValueRange
ScpChGetDataValueMin
ScpChGetDataValueMax
Since
0.4.4