LibTiePie
0.4.2
Library for interfacing TiePie engineering instruments
|
This section contains all trigger hold off related functions. More...
Functions | |
uint64_t | ScpGetTriggerHoldOffCountMax (TpDeviceHandle_t hDevice) |
Get maximum trigger hold off count in samples. | |
uint64_t | ScpGetTriggerHoldOffCount (TpDeviceHandle_t hDevice) |
Get trigger hold off count in samples. | |
uint64_t | ScpSetTriggerHoldOffCount (TpDeviceHandle_t hDevice, uint64_t qwTriggerHoldOffCount) |
Set trigger hold off count in samples. | |
This section contains all trigger hold off related functions.
Trigger hold off defines a period of time after starting a measurement in which the oscilloscope will not respond to triggers. It is used to get stable triggering on periodical signals that have multiple moments that would otherwise generate a trigger.
When pre samples are selected, trigger hold off can be used to ensure that all pre samples are measured before a trigger occurs. Set the trigger hold off to TH_ALLPRESAMPLES to measure all pre samples and have no invalid pre samples.
Trigger hold off is set in a number of samples. By default the trigger hold off is set to: 0 samples.
uint64_t ScpGetTriggerHoldOffCount | ( | TpDeviceHandle_t | hDevice | ) |
Get trigger hold off count in samples.
hDevice | A device handle. |
uint64_t ScpGetTriggerHoldOffCountMax | ( | TpDeviceHandle_t | hDevice | ) |
Get maximum trigger hold off count in samples.
hDevice | A device handle. |
uint64_t ScpSetTriggerHoldOffCount | ( | TpDeviceHandle_t | hDevice, |
uint64_t | qwTriggerHoldOffCount | ||
) |
Set trigger hold off count in samples.
hDevice | A device handle. |
qwTriggerHoldOffCount | Trigger hold off count in samples. |